The HR4000CG-UV-NIR spectrometer is perfect for applications where fast measurements and high resolution are needed.
- Flexible – programmable microcontroller
- Compatible – triggering functions synchronize the spectrometer to other devices
Sales, Service
& Support Facility
Duiven, The Netherlands
Ocean Optics GmbH Sales,
Service & Support Facility
Ostfildern, Germany
$6,563.00
This product is currently out of stock and unavailable.
The HR4000CG-UV-NIR spectrometer is perfect for applications where fast measurements and high resolution are needed.
Engineering Specifications | HR4000CG-UV-NIR |
PHYSICAL | |
Dimensions: (L x W x H) mm and inches | 148.6 x 104.8 x 45.1 mm (5.9 x 4.1 x 1.8 in.) |
Weight: kg and lb | 0.57 kg (1.26 lbs.) |
DETECTOR | |
Type: | Toshiba TCD1304AP |
Range: | 190 – 1100 nm |
Quantum efficiency: | no |
SPECTROSCOPIC | |
Wavelength range: | 200 – 1100 nm |
Integration time: | 4 milliseconds to 20 seconds (continuous); 10 microseconds to 4 milliseconds (shutter) |
Dynamic range: | 2 X 109 (system); 2000:1 for a single acquisition |
Signal-to-noise ratio: | 300:1 (at full signal) |
Grating: | HC-1, 300 lines per mm grating |
Slit: | 5 µm wide |
Optical resolution: | <1.0 nm FWHM (note: resolution varies at either end of the spectral range) |
Stray light: | <0.05% at 600 nm; <0.10% at 435 nm |
Buffering: | no |
Fiber optic connector: | SMA 905 to single-strand optical fiber (0.22 NA) |
ELECTRONICS | |
Power consumption: | 450 mA @ 5 VDC |
Strobe functions: | 2 programmable strobe signals (single/continuous) |
Interfaces: | USB 2.0, 480 Mbps; 2-wire RS-232; I2C 2-wire serial bus |
Temperature: | -30° to 70° C Storage & -10° to 50° C Operation |
Humidity: | 0% – 90% non-condensing |
Applications in which this product is used:
Measurement Techniques in which this product is used: